PRIMING THE INFUSION LINE
SECTION 14
The priming function allows for lling the infusion line with
the drug contained in the reservoir.
The volume available for priming is 1.5 ml.
The priming function is enabled when the device is
switched on and the pusher is in the infusion start position,
regardless of whether the settings lock is on.
The priming procedure is as follows:
1. switch on the device as described in the relevant section;
2. the display shows Pr. There are three options:
a. postpone the priming;
b. cancel the priming;
c. perform the priming.
a. Postpone the priming
Press the
and buttons together: the pump switches
off and the display shows OFF. Or wait 10 seconds and the
pump switches itself off automatically.
b. Cancel the priming
Press the
button; the pump starts the infusion and the
display shows the time remaining until the end of the infu-
sion, as well as the current ow rate.