Cane Crono Guide de l'utilisateur Page 29

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 114
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 28
29
MAN 01/EN/03 CRONO PAR 50 06/13
POST-OCCLUSION BOLUS
WARNINGS
SECTION 7
INFUSION SET OCCLUSION
The pump is designed to recognise when the administration
of a drug has been interrupted by external means, such as,
for example, the kinking of the infusion set tube and the
consequent occlusion.
In these circumstances, the pump stops the infusion: the
display indicates that there is an occlusion, accompanied
by a brief beep. The pump then continues beeping every
10 seconds.
While the system is still occluded, the drug is not
administered: to recommence the infusion, press the
button after having removed the cause of the occlusion.
The cause of the occlusion is to be found along the infusion line and at the
point of injection.
To avoid or reduce the incidence of occlusions, you are advised to use an
infusion set with anti-kinking tubes.
NOTES
The occlusion alarm is given when the pump detects excessive back pressure
in the infusion line. This back pressure must be removed without releasing a
post-occlusion bolus, which could cause serious harm to the patient.
The volume of a CRONO PAR 50, post-occlusion bolus, considering only the
combined volume of the pump and the reservoir is about 0.9 ml (in PL2 mode).
The volume of the bolus released after an occlusion can vary, depending on
the infusion set and all the other components that comprise the infusion line.
Another element that could affect the volume of the released bolus after an
occlusion is the presence of any air in the system.
After the occlusion alarm is given, take any and all measures appropriate
to avoid the administration of a post-occlusion bolus, such as, for example,
disconnecting the infusion set from the patient.
Vue de la page 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 113 114

Commentaires sur ces manuels

Pas de commentaire