Cane Crono Guide de l'utilisateur Page 69

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 114
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 68
69
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SECTION 14
SWITCHING ON THE PUMP
From the OFF condition, press the button. The pump
will give a brief beep and display:
Pr (priming function)
There are three options:
a. postpone the priming;
b. cancel the priming;
c. perform the priming.
Switching on the pump
After you have performed the priming, or if you switch the
pump on when it is in the StoP condition, the display shows
the infusion time in hours or in hours and/or minutes or just
minutes.
Press the button again, and the display shows the
corresponding hourly ow value.
During the infusion the time shown on the display is counted down:
hour by hour if the time shown on the display is expressed in hours, with a
ashing letter h;
minute by minute if the time shown on the display is expressed in hours
and minutes or just minutes.
Before starting an infusion:
check that the reservoir is correctly connected;
Inspect the infusion line to ensure there are no folds, clamps or other
occlusions in the line;
expel any air bubbles;
in case different hourly ows are used, check that the time is correctly set.
NOTE
WARNINGS
ml/h
Vue de la page 68
1 2 ... 64 65 66 67 68 69 70 71 72 73 74 ... 113 114

Commentaires sur ces manuels

Pas de commentaire