MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SETTING THE BOLUS DOSE VOLUME
Proceed as follows:
1. the letter d is shown on the secondary display and
the volume of the bolus dose in the primary display. The
numeric values are ashing but the d is steady;
2. press the
button to increase the value; press the
button to decrease it. Each change is indicated by a beep;
3. if you do not press any button for 20 seconds, the settings
session is terminated and the message OFF or StoP is
displayed;
4. press the
button before the OFF or StoP indication
is displayed (while the bolus dose volume is still ashing)
to continue to the setting of the next parameter: SETTING
THE INTERVAL BETWEEN BOLUS DOSES.
If you keep pressing either the or buttons, it is possible to quickly
change the bolus dose volume.
The bolus dose setting is automatically stored in the pump’s memory.
If you press the
button when the display indicates d0,02, the value chan-
ges to d0,00: this setting disables the bolus dose function.
NOTES
SECTION 14
PROG
PROG
If the bolus dose volume is set to 0.00 ml, pressing the button during the
infusion will display the Er, d error message.
WARNING
The bolus dose volume can be set from 0.00 to 2.00 ml in increments of 0.022
ml.