SETTING THE INTERVAL BETWEEN BOLUS DOSES
The interval between bolus doses can be set to a value
between no,Lt (function disabled) and 24 hours, in the
following ways:
from 5 minutes to 1 hour in increments of 5 minutes;
from 1 hour to 24 hours in increments of 15 minutes.
This applies a temporal limit to the bolus dose function.
Proceed as follows:
1. the display shows the ashing value of the bolus dose
interval;
2. press the button to increase the value; press the
button to decrease it. Each change is indicated by a beep;
3. if you do not press any button for 20 seconds, the settings
session is terminated and the message OFF or StoP is
displayed;
4. press the button before the OFF or StoP indication
is displayed (while the bolus dose interval is still ashing)
to continue to the setting of the next parameter: SETTING
THE FIRST FLOW RATE (F1).
If you keep either the or button pressed, it is possible to quickly change
the interval between the bolus doses.
The bolus dose interval setting is automatically stored in the pump's memory.
Press the button with the display showing it 5’, the display changes to
no,Lt: this setting disables the function that regulates the interval between
bolus doses, and thus disables any time restrictions on the administration of
bolus doses.
NOTES
PROG
PROG
PROG
SECTION 11
PROG
ml/h