DISPLAYING THE NUMBER OF ADMINISTERED BOLUS DOSES
DURING AN INFUSION AND THE NUMBER OF INFUSIONS
1 - To check the number of bolus doses administered by
the patient during an infusion while the pump is in OFF or
StoP, press and hold the
button for a few seconds and
the value will be displayed.
2 - To check the number of infusions delivered by the device
while it is OFF or StoP, press and hold the
button for
about 8 seconds.
The number of the administered bolus doses is only available until the begin-
ning of a new infusion; switching the pump on for a new infusion automatically
resets the value.
The number of bolus doses includes those interrupted by pressing the
button.
WARNINGS
RESTRICTIONS ON THE ADMINISTRATION OF BOLUS DOSES
If the bolus dose volume is set to 0.00 ml, pressing the
button will display the Er,d error message.
The administration of bolus doses is limited by the interval
between doses (temporal limitation).
By setting no,Lt you can disable this limitation.
Bolus doses cannot be administered during the interval
period observed by the function.
During the limitation period, if you request a bolus dose,
the display shows Er,t (temporal limitation between active
bolus doses).