MAN 01/EN/03 CRONO PAR 50 06/13
DISPLAY
AUDIBLE
SIGNAL
ERROR
DESCRIPTION
CORRECTIVE
ACTION
Beep repeated every
10 sec approximately.
When a battery is inserted
and at the start of every
infusion, the device runs
a control algorithm on
the parameters stored in
the memory. If an error is
found, the manufacturer
default settings are
restored, the motor stops
and an error is shown on
the display.
Initialise the device
(see page 30) and
set the prescribed
infusion parameters.
Beep repeated every
10 sec approximately.
Anomaly in the safety
circuit which drives the
pump motor. If an error is
found, the pump locks and
the error is indicated.
Initialise the device
(see page 30).
Beep repeated every
10 sec approximately.
Anomaly in the pusher
system.
Initialise the device
(see page 30).
Beep repeated every
10 sec approximately.
Mechanism blocked
because of an occlusion in
the infusion line.
Eliminate the cause
and press the
button
(see page 29).
Beep when you press
the button.
The bolus dose has been
set to 0.00 ml.
This function is not
available.
(see pages 50 and
73).
Beep when you press
the
button.
The bolus dose cannot be
administered because the
dose interval imposed by
the function has not been
completed.
This function is not
available.
(see pages 50 and
73).